To continually enhance the management technique by virtue of your rule of "sincerely, great faith and high-quality are the base of company development", we widely absorb the essence of similar merchandise internationally, and continuously build new merchandise to meet the demands of customers for Dc Input Power Supply, Lcr Meter Inductance, Dc Tester, We're going to endeavor to maintain our fantastic track record as the very best products supplier in the planet. When you've got any questions or reviews, you should get in touch with with us freely.
Chinese wholesale Loop Calibrator 4-20ma – ET2110 High-Precision Temperature Process Calibrator – Zhongchuang Detail:

1

¤ Measurement/output: voltage, millivolt, resistance, resistance, thermocouple, rtd, current, etc

¤ 220V measurement function

¤ Simulating 2-wire transmitter

¤ Resistance measurement options: 2 wires, 3 wires, 4 wires

¤ Accuracy: 0.01%, 0.02%

¤ Two isolated channels support measure and output simultaneously

¤ Providing manual step, automatic step, automatic step and manual step functions.

¤ 3.5 TFT LCD screen,resolution rate 480*320

¤ Measurement and output data can be displayed simultaneously or separately;

¤ 5000mAh Lithium battery

¤ Automatic power shutdown function, shutdown time can be set, and suitable for on-site use

¤ Providing DC 24V loop power for on -site use

¤ Thermocouple measurement and output provide three kinds of cold junction temperature compensation methods: built-in, external and manual, among them, external reference junction adopts A class Pt100 Platinum resistance, which can correct temperature through inputting certificate value

¤ Thermocouple types: R,S,K,E,J,T,N,B,L,U,XK,WRE325,WRE526;

¤ RTDtypes:PT100-385,PT100-392,PT100-JIS,PT200-385,PT500-385,PT1000-385,Cu10,Cu50,Cu100,Nil20,BA1,BA2,PT10;

¤ Temperature difference measurement function: the accuracy is up to 0.003℃. This product can measure the temperature difference between two points in the space, and complete a temperature difference data collection within 0.4 seconds, effectively improving the measurement accuracy.The 10-minute fluctuation during the test can be calculated in real time.Before using standard platinum resistance or standard thermocouple for measuring work, the measurement results can be traced through the input of certificate value, and meet the requirements of the thermostatic tank test specification for electrical measuring instruments.

¤ Standard temperature measurement function: Comparing with common thermocouple measurement and RTD measurement,the difference is that  this measurement method can trace the temperature by certificate value,the supported standard thremocouple and RTD are as followed: S、R、B、T, Pt25、Pt100 

¤ Arbitrary sensor’s measurement function; Users can transform the measured physical quantity(Pressure, flow speed, temperature,etc) to voltage, current, resistance,etc conveniently for measurement. Besides, Users only need to input the response curve in advance, and the multimeter will adopt the internal algorithm for numerical conversion and correction, and finally the measured physical quantity will be displayed on the screen.You are free to edit and modify the display units of the measured physical quantities.

¤ Precision temperature control function; Precision temperature control function will realize the temperature closed-loop control of thermostatic equipment, which replace the high precision PID controller.Under the condition of thermostatic equipment and network voltage,temperature fluctuation will be better than 0.02℃/10min (Thermostatic Bath)

¤ Measurement function of ρ’s value: can measure duty ratio of periodic square wave signal; verify and calibrate the PID parameter of Various digital temperature indicating regulators which is outputted by time scale,and conform to the requirement of 《JJG617-1996 digital temperature indicating regulators》

¤ Thermal Conversion Function: realize the conversion between various electric quantity and temperature.The types of electric quantity and temperature conversion include: working thermocouple, industrial rtd and various temperature transmitter.

¤ Numerical setting mode:its with the most flexible and convenient way to set output value; User can use the numerical keyboard to set output value directly, and can realize the incremental setting by direction key. In addition, the equipment also has a step or ramp numerical setting mode that can be numbered.

¤ Sinusoidal output function: The verification/calibration of some process logger (especially mechanical logger); Usually It involves running test, and it can provide signals to the measured table by using sinusoidal output mode.

¤ Data Record Function:  with powerful record management function, it can establish up to 32 device numbers. Each device number has 16 record pages, and each record page contains four basic information: time, measured value, output value and custom value.Users can carry out equipment management, record deletion and other operations according to requirements.

      Model

Accuracy

Temp Range

Optional Function

ET2110B

0.01级

15~25℃

For optional function,please contact us for detail code information about its relevant function

ET2110C

0.02级

ET2110BT

0.01级

0~50℃

ET2110CT

0.02级

Function

Range

Resolution rate(0.01%)

Resolution rate(0.02%)

Accuracy(0.01%)

Accuracy(0.02%)

Note

DC Output

Voltage

100mV

0.1µV

1µV

0.005%+0.003%

0.01%+0.005%

Max load current    <=2.5mA

1V

1µV

10µV

0.005%+0.001%

0.01%+0.005%

10V

10µV

100µV

0.005%+0.001%

0.01%+0.005%

Current (Active/Passive)

30mA

0.1µA

1µA

0.005%+0.003%

0.01%+0.003%

Max load voltage(active output)   20V

Resistance

50Ω

0.1mΩ

0.005%+10 mΩ

0.01%+15 mΩ

Excitation current 0.4-4mA

500Ω

1mΩ

0.005%+20 mΩ

0.01%+30 mΩ

Excitation current  0.1-2mA

5000Ω

10mΩ

0.005%+50 mΩ

0.01%+50 mΩ

Excitation current 0.04-0.4mA

24V

24V

 

±10%

Loop output

RTD

See detail of RTD sheet

 

Thermocouple

See detail of thermocouple sheet

 

DC Measurement

Voltage

200mV

0.1µV

0.005%+0.003%

0.01%+0.005%

 

2V

1µV

0.005%+0.001%

0.01%+0.005%

 

20V

10µV

0.005%+0.001%

0.01%+0.005%

 

200V

100µV

0.005%+0.001%

0.01%+0.005%

 

Current

20mA

0.1µA

0.005%+0.003%

0.01%+0.003%

 

200mA

1µA

0.005%+0.003%

0.01%+0.003%

 

Resistance (4-wire)

50Ω

0.1mΩ

0.005%+10 mΩ

0.01%+15 mΩ

Excitation current  1mA

500Ω

1mΩ

0.005%+20 mΩ

0.01%+30 mΩ

5kΩ

10mΩ

0.005%+50 mΩ

0.01%+50 mΩ

Excitation current  0.1mA

Resistance (2,3-wire)

50Ω

0.1mΩ

0.005%+30 mΩ

(三线)

0.005%+50 mΩ

(二线)

0.005%+35 mΩ

(三线)

0.005%+60 mΩ

(二线)

Excitation current  1mA

500Ω

1mΩ

5kΩ

10mΩ

0.005%+80mΩ

0.01%+80 mΩ

Excitation current  0.1mA

RTD

See detail of RTD sheet

 

Thermocouple

See detail of thermocouple sheet

 

AC Measurement

AC Voltage

200mV

1µV

±(0.2%+100)(40Hz-30kHz)

 

2V

10µV

 

20V

100µV

±(0.2%+100)(40Hz-5kHz)

±(0.8%+300)(5k-30kHz)

 

200V

1mV

±(0.2%+200)(40Hz-5kHz)

±(0.8%+450)(5k-30kHz)

 

AC Current

20mA

0.1µA

±(0.3%+400)(40Hz-5kHz)

 

200mA

1µA

 

RTD Sheet

Signal Types

Temp Range

Resolution rate

Accuracy (0.01%)

Accuracy (0.02%)

Note

PT10

-200-850℃

0.01℃

0.1℃

0.2℃

4 -wire measurement

PT100-385

-200-850℃

0.01℃

0.1℃

0.2℃

PT100-392

-200-850℃

0.01℃

0.1℃

0.2℃

PT100-JIS

-200-850℃

0.01℃

0.1℃

0.2℃

PT200-385

-200-630℃

0.01℃

0.1℃

0.2℃

PT500-385

-200-630℃

0.01℃

0.2℃

0.3℃

PT1000-385

-200-650℃

0.01℃

0.1℃

0.2℃

Cu10

-100-260℃

0.01℃

0.5℃

0.6℃

Cu50

-50-150℃

0.01℃

0.15℃

0.25℃

Cu100

-50-150℃

0.01℃

0.08℃

0.2℃

BA1

-200-650℃

0.01℃

0.4℃

0.5℃

BA2

-200-650℃

0.01℃

0.25℃

0.3℃

Ni20

-80-260℃

0.01℃

0.3℃

0.4℃

Thermocouple Sheet

Signal Types

Temp Range

Resolution rate

Accuracy (0.01%)

Accuracy (0.02%)

Note

K

-200-0℃

0-1372℃

0.1℃

0.4℃

0.3℃

Exclude accuracy of cold junction compensation

R

-50-0℃

0-1768℃

0.1℃

0.9℃

0.7℃

S

-50-0℃

0-1768℃

0.1℃

0.9℃

0.6℃

E

-50-0℃

0-1000℃

0.1℃

0.5℃

0.4℃

J

-200-0℃

0-1200℃

0.1℃

0.2℃

0.1℃

T

-100-0℃

0-400℃

0.1℃

0.3℃

0.15℃

L

-200-900℃

0.1℃

0.2℃

N

-200-0℃

0-1300℃

0.1℃

0.3℃

0.2℃

B

600-1820℃

0.1℃

0.6℃

U

-200-0℃

0-400℃

0.1℃

0.4℃

0.2℃

XK

-200-800℃

0.1℃

0.5℃

WRE325

0-1500℃

0.1℃

0.5℃

WRE526

0-1500℃

0.1℃

0.4℃

Note: accuracy of built-in cold junction compensation is 0.5℃


Product detail pictures:

Chinese wholesale Loop Calibrator 4-20ma – ET2110 High-Precision Temperature Process Calibrator – Zhongchuang detail pictures


Related Product Guide:

Our well-equipped facilities and great excellent command throughout all stages of generation enables us to guarantee total customer fulfillment for Chinese wholesale Loop Calibrator 4-20ma – ET2110 High-Precision Temperature Process Calibrator – Zhongchuang , The product will supply to all over the world, such as: Congo, Cologne, Armenia, Over the years, with high-quality solutions, first- class service, ultra-low prices we win thee trust and favor of customers. Nowadays our products sell all over the domestic and abroad. Thanks for the regular and new customers support. We offer high quality product and competitive price, welcome the regular and new customers cooperate with us!
  • Although we are a small company, we are also respected. Reliable quality, sincere service and good credit, we are honored to be able to work with you!
    5 Stars By Ellen from kazan - 2018.06.26 19:27
    Superb technology, perfect after-sales service and efficient work efficiency, we think this is our best choice.
    5 Stars By Jodie from Slovakia - 2018.09.29 17:23
    Write your message here and send it to us