We attempt for excellence, provider the customers", hopes to be the most beneficial cooperation team and dominator enterprise for staff, suppliers and shoppers, realizes value share and continuous advertising for Pressure Calibrator, Blackbody Furnace For Thermometer Calibration, Multifunction Process Calibrator, We welcome new and previous customers from all walks of lifestyle to call us for upcoming business enterprise interactions and reach mutual accomplishment!
OEM/ODM China High Frequency Lcr Meter - ET1090/ET1091 Benchtop LCR Meter for Component Measurement – Zhongchuang Detail:

¤ The highest measuring frequency is 100kHz, and the stepping; frequency is continuously adjustable at 1Hz;

¤ Test level 10~2000 mV, 1 mV step continuously adjustable;

¤ Supporting DC Resistance (DCR) and Electrolytic Capacitance Measurement;

¤ Internal bias voltage output 10mV~1500mV;

¤ 3.5-inch TFT display, 5-bit display;

¤ Data recording function (maximum, minimum, average);

¤ Supporting SCPI Communication Protocol;

¤ Provide system settings to configure Chinese or English, buzzer, screen brightness, etc;

¤ The basic measurement accuracy is 0.2%;

¤ Manual and automatic range;

¤ With open circuit and short circuit calibration function;

¤ Comparator sorting (5 gears), alarm function. 

Model

ET1090A

ET1090B

ET1090C

ET1091A

ET1091B

ET1091C

Testing frequency

10 points (100, 120, 200, 400, 800, 1K, 2K, 4K, 8K, 10K)

12 points (100, 120, 200, 400, 800, 1K, 2K, 4K, 8K, 10K, 15K, 20K)

16 points (100, 120, 200, 400, 800, 1K, 2K, 4K, 8K, 10K, 15K, 20K, 40K, 50K, 80K, 100K)

10-10KHz continuous adjustable, 1Hz stepby-step

10-10KHz

continuous

adjustable, 1Hz stepby-step

10-100KHz

continuous

adjustable, 1Hz step by-step

Display screen

Display screen 3.5 inch TFT LCD

Number of display digits

Principal parameter: 5 digit,Secondary parameter: 5 digit

Measured parameter

Principal parameter: L/C/R/Z,Secondary parameter: X/D/Q/θ/ESR

Measurement Range

L:0.001μH~9999H

C:0.001pF~99.999mF

R:0.0001Ω~99.99MΩ

Basic accuracy

0.20%

Measuring display speed

2 time/s (slow), 4 times/s (medium), 8 time/s (fast)

Internal bias

0-1500mV adjustable, at a step of 1mV.

Testing level

Six fixed level(0.1V、0.3V、0.6V、1V、1.5V、2V)

0.1~2V adjustable, at a step of 1mV.

Signal source output

impedance

30Ω、100Ω

Calibration function

Open circuit calibration, short circuit calibration

Screening function

The limit range of screening can be set to -50%~+50%, and the fixed points are 1%, 5%, 10% and 20%.

Comparator

5 groups sorting,3 groups of qualified setting, one group of unqualified setting, one group of auxiliary setting

Interfaces

standard: RS232(or 485), USB Device, Handler ;optional: GPIB,USB Host

Others

Support dc resistance (DCR), electrolytic capacitor measurement model,Adjustment of backlight brightness, Chinese and English are optional

¤ Power supply voltage: 220V.AC±10%, or 110V.AC±10%, 45~65Hz;

¤ Power consumption: <10W;

¤ Display: 3.5 inch TFT LCD screen, resolution 480 *320, color 16M;

¤ Temperature range: Operating state 10℃~+40℃, non-operating state -10℃ ~+60℃;

¤ Humidity range: 0 ~ 40 C, < 90% relative humidity;

¤ Interface: RS232 (or 485), USB Device, Handle, GPIB, USB Host. 

¤ Three core power cord(30A51)

¤ 4-terminals of kelvin testing cable(35A51) 

ET1090ET1091 Benchtop LCR Meter for Component Measurement 1
ET1090ET1091 Benchtop LCR Meter for Component Measurement 2
ET1090ET1091 Benchtop LCR Meter for Component Measurement 3
ET1090ET1091 Benchtop LCR Meter for Component Measurement 4


Product detail pictures:

OEM/ODM China High Frequency Lcr Meter - ET1090/ET1091 Benchtop LCR Meter for Component Measurement – Zhongchuang detail pictures

OEM/ODM China High Frequency Lcr Meter - ET1090/ET1091 Benchtop LCR Meter for Component Measurement – Zhongchuang detail pictures

OEM/ODM China High Frequency Lcr Meter - ET1090/ET1091 Benchtop LCR Meter for Component Measurement – Zhongchuang detail pictures

OEM/ODM China High Frequency Lcr Meter - ET1090/ET1091 Benchtop LCR Meter for Component Measurement – Zhongchuang detail pictures


Related Product Guide:

We are commitment to offer the competitive price ,outstanding products quality, as well as fast delivery for OEM/ODM China High Frequency Lcr Meter - ET1090/ET1091 Benchtop LCR Meter for Component Measurement – Zhongchuang , The product will supply to all over the world, such as: Pakistan, Riyadh, Paraguay, Based on our automatic production line, steady material purchase channel and quick subcontract systems have been built in mainland China to meet customer's wider and higher requirement in recent years. We have been looking forward to cooperating with more clients worldwide for common development and mutual benefit!Your trust and approval are the best reward for our efforts. Keeping honest, innovative and efficient, we sincerely expect that we can be business partners to create our brilliant future!
  • Wide range, good quality, reasonable prices and good service, advanced equipment, excellent talents and continuously strengthened technology forces,a nice business partner.
    5 Stars By Emma from Moscow - 2017.10.23 10:29
    The factory technical staff not only have high level of technology, their English level is also very good, this is a great help to technology communication.
    5 Stars By James Brown from Netherlands - 2017.11.11 11:41
    Write your message here and send it to us